Switching Setup Error
Cause: This error indicates a potential problem with the setup of the complete circuit of the switching system. This most frequently occurs in DIT-MCO wiring analyzers with multiple switching modules. If the cabling is incorrect, not continuous, or there is a fault in the switching, the test operator may receive this error message.
Solution: Steps to resolve the matter include:
- Check the cabling to the switching modules for loosening or damage, and then verify that the mating connectors are fully and firmly seated.
- The Distributed Switching Configuration file is set correctly for the amount of switching the system contains. This applies only to Model 26XX wiring analyzers.
- The system's switching is contiguous, with no gaps between modules. For example, if a user has five modules of switching and, for some reason, does not want to use module 4, the user cannot use modules 1-3 and 5. Module 5 would have to be reassigned to module 4 to operate correctly.
Source: NETS User Manual, Revision D, document 132-06035-0006, section 11.2.7.